Mauricio Vergara E(2), Drs. Marcelo Castro S(1), José Matas N(1),
Mauricio Arias B(1,2), Gumaro Martínez C(1), Julio Rosales L(1).
1.
Departmentos of Radiología y Medicina, Hospital Militar
2. Departmento
de Radiología, VidaIntegra. Santiago de Chile.
Abstract
Tomography
studies are less than 10% of the total number of imaging studies but deliver
more than 40% of the total collective dose to those patients. This situation
has generated an increasing number of initiatives to reduce tomography
doses. We analysed our routine CT protocols for headache, abdomen, patellofemoral
dysfunction and sinusitis; they conform 60% of our total demand for CT
examinations in a hospital setting. This revision enabled us to reduce
the dose the patient in 30 to 50% of their original values. Diagnostic
quality was maintained and the benefit to the patient is obvious.
Key
words: Computed tomography, Radiation.
Resumen
Los estudios
de tomografía computada corresponden al 10% del total de los estudios
radiológicos pero superan el 40% de la dosis de radiación
colectiva al paciente. Ello ha generado una gran presión internacional
por reducir las dosis en tomografía computada. Se revisaron los
protocolos vigentes en estudio de cefalea, abdomen, disfunción
patelo-femoral y sinusitis. Ellos conforman aproximadamente el 60 % de
los estudios de tomografía computada en un centro hospitalario.
Esta revisión, tanto bibliográfica como de series técnicas
propias, permitieron reducir las dosis de radiación al paciente
en un 30% a un 50% de sus valores originales. Ello sin disminuir su calidad
diagnóstica y con obvio beneficio para el paciente.
Palabras
clave: Radiación, Tomografía computada.
 |
|
|
| |
|
|
Vergara
M. Reducción de dosis al paciente en TC: Un estudio de resultados
en la práctica clínica. Rev Chil Radiol 2003; 9:29- 32.
.